The Keysight Technologies U2001H is a high-performance, high-power USB power sensor designed for fast, accurate power measurements. Operating across a frequency range of 10 MHz to 6 GHz with a power measurement range from -50 dBm to +30 dBm, this sensor eliminates the need for a separate power meter by connecting directly to a PC or compatible Keysight instrument via a USB 2.0 interface. Its high damage threshold and excellent VSWR make it highly reliable for RF and microwave power testing applications.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 (USBTMC-USB488) |
| Display | None (requires host PC or instrument) |
Physical
| Parameter | Value |
|---|---|
| Operating Temperature | 0 to 55 °C |
| Dimensions (W×H×D) | 46 mm x 43 mm x 164 mm |
RF Input and Measurement Performance
| Parameter | Value |
|---|---|
| Frequency Range | 10 MHz to 6 GHz |
| Power Measurement Range | -50 dBm to +30 dBm |
| Input Impedance | 50 Ω |
| RF Input Connector Type | N-type (male) |
Connectivity and Control Interfaces
| Parameter | Value |
|---|---|
| Host Interface Type | USB 2.0 |
| USB Class Type | USBTMC-USB488 |
| External Trigger Input Connector | BNC (via optional trigger cable) |
| External Trigger Output Connector | BNC (via optional trigger cable) |
Software and Compatibility
| Parameter | Value |
|---|---|
| Supported Software Applications | Keysight BenchVue, BV0007B Power Meter App |
| Driver Availability | IVI-C, IVI-COM, LabVIEW |
| Command Set Compatibility | SCPI |
Power and Electrical Requirements
| Parameter | Value |
|---|---|
| Power Source | USB Port |
Environmental Specifications
| Parameter | Value |
|---|---|
| Operating Temperature Range | 0 to 55 °C |
| Storage Temperature Range | -40 to 70 °C |
| Operating Humidity | 95% RH at 40 °C (non-condensing) |
| Storage Humidity | 90% RH at 65 °C (non-condensing) |
| Operating Altitude | Up to 4600 m (15000 ft) |
| Storage Altitude | Up to 15300 m (50000 ft) |
Regulatory and Standards Compliance
| Parameter | Value |
|---|---|
| Recommended Calibration Cycle | 1 year |
Performance
| Parameter | Value |
|---|---|
| Internal Zeroing and Calibration | Supported, eliminates the need for an external reference source |
Measurement Performance
| Parameter | Value |
|---|---|
| Averaging Range | 1 to 1024 (manual or auto) |
Technology
Diode Power Sensor
Applications
RF & Microwave Power Measurement