The Keysight W8486A (formerly Agilent) is a waveguide diode power sensor designed for precise average power measurements in the W-band frequency range from 75 GHz to 110 GHz. Featuring an exceptionally low SWR of 1.08 to minimize mismatch uncertainty, it covers a dynamic range from -30 dBm to +20 dBm. It is compatible with a wide array of legacy and modern Keysight power meters and comes with a 50 MHz calibration port with calibration traceability to NIST.

Specifications

General

ParameterValue
InterfacesProprietary 12-pin connector for sensor cable
DisplayNone (requires external power meter)
Supplied AccessoriesW281C waveguide to coax adapter (50 MHz calibration adapter)

Physical

ParameterValue
Power SupplyPowered by compatible power meter via sensor cable
Operating Temperature0 to 55 °C
Dimensions (W×H×D)41 mm × 35 mm × 121 mm
Weight0.17 kg
Shipping Weight0.64 kg

Measurement & Performance

ParameterValue
Frequency Range75 GHz to 110 GHz
Power Measurement Range-30 dBm to +20 dBm
Voltage Standing Wave Ratio (VSWR)1.08
Sensor Technology TypeDiode
Measurement TypeAverage power

System Interface & Connectivity

ParameterValue
RF Waveguide Port TypeUG-387/U waveguide flange
Waveguide Size DesignationWR-10
Compatible Power Meter ModelsEPM, EPM-P, P-series, E1416A VXI, 70100A, and 43X series

Calibration & Accuracy

ParameterValue
Calibration Port50 MHz
TraceabilityUS National Institute of Standards and Technology (NIST) millimeter-wave sensor calibration

Environmental

ParameterValue
Storage Temperature Range-40 °C to +75 °C
Operating Humidity95% R.H. at 40 °C (non-condensing)

Power Limits

ParameterValue
Maximum Input Power100 mW (+20 dBm) average/peak

Connectivity

ParameterValue
Compatible Sensor CablesKeysight 11730 series

Performance

ParameterValue
Zero Set±200 nW
Zero Drift±40 nW (over 1 hour at constant temperature)
Measurement Noise< 80 nW
Power Linearity-30 to +15 dBm: ±1%, +15 to +20 dBm: +2% / -4%

Inputs & Outputs

ParameterValue
Waveguide FlangeUG-387/U-M circular flange

Product Options

ParameterValue
Option 100Excludes the 50 MHz Type-N calibration adapter

Technology

Diode

Applications

RF and Microwave Power Measurement