The Keysight U8488A is a high-performance USB thermocouple power sensor designed for fast, accurate, and reliable RF and microwave power measurements from 10 MHz to 67 GHz without the need for external calibration.

Specifications

General

ParameterValue
InterfacesUSB 2.0 (USBTMC compliant)
DisplayNone (PC or instrument-based display via software)
Sensor TechnologyThermocouple
Warm-up Time30 minutes
Recommended Calibration Cycle1 year

Physical

ParameterValue
Power SupplyUSB Bus Powered (5 V, < 500 mA)
Operating Temperature0 to 55 °C
Dimensions (W×H×D)53 mm × 35 mm × 123 mm
Weight0.222 kg

RF and Microwave Performance

ParameterValue
Frequency Range10 MHz to 67 GHz
Power Range-35 dBm to +20 dBm
Maximum Average Power (Damage Level)+25 dBm
Maximum Peak Power (Damage Level)+30 dBm (< 10 µs)
Maximum DC Voltage15 V
RF Connector Type1.85 mm male
Input Impedance50 Ω
SWR (10 MHz to 100 MHz)≤ 1.15
SWR (100 MHz to 2 GHz)≤ 1.15
SWR (2 GHz to 12.4 GHz)≤ 1.18
SWR (12.4 GHz to 26.5 GHz)≤ 1.22
SWR (26.5 GHz to 40 GHz)≤ 1.30
SWR (40 GHz to 50 GHz)≤ 1.35
SWR (50 GHz to 67 GHz)≤ 1.41

Metrological and Calibration Specifications

ParameterValue
Internal Zero and Calibration SupportYes, built-in (eliminates need for external calibration source)

Dynamic and Measurement Speed Parameters

ParameterValue
Measurement Speed (Fast Mode)Up to 900 readings/s
Measurement Speed (Normal Mode)Up to 40 readings/s

Software and Programming Support

ParameterValue
Standard Programming LanguageSCPI
Supported SoftwareKeysight BenchVue, BV0007B Power Meter App

Environmental and Safety Compliance

ParameterValue
Storage Temperature-40 °C to 70 °C
Operating HumidityUp to 95% RH at 40 °C (non-condensing)
Storage HumidityUp to 90% RH at 65 °C (non-condensing)
Operating AltitudeUp to 3,000 m (9,840 ft)
Storage AltitudeUp to 15,420 m (50,000 ft)

Performance

ParameterValue
Zero Set< 25 nW
Zero Drift< 7 nW (over 1 hour at constant temperature)
Measurement Noise< 80 nW (free run, 1 average)
Power Linearity±0.80% (-35 to +20 dBm, 25 °C ± 5 °C)
Calibration Factor Uncertainty (10 to 100 MHz)1.6% (25 °C ± 5 °C)
Calibration Factor Uncertainty (100 MHz to 2 GHz)1.4% (25 °C ± 5 °C)
Calibration Factor Uncertainty (2 to 12.4 GHz)1.6% (25 °C ± 5 °C)
Calibration Factor Uncertainty (12.4 to 18 GHz)1.7% (25 °C ± 5 °C)
Calibration Factor Uncertainty (18 to 26.5 GHz)2.0% (25 °C ± 5 °C)
Calibration Factor Uncertainty (26.5 to 40 GHz)2.1% (25 °C ± 5 °C)
Calibration Factor Uncertainty (40 to 50 GHz)2.4% (25 °C ± 5 °C)
Calibration Factor Uncertainty (50 to 67 GHz)3.2% (25 °C ± 5 °C)

Power

ParameterValue
Current Consumption< 420 mA (operating)

Frequency Options

ParameterValue
Low-frequency Limit (Option CFT)100 kHz

Triggering

ParameterValue
Trigger SourceFree run, Internal, External, Software
Trigger Delay Range1 µs to 10 s
Trigger Delay Resolution1 µs
Trigger Jitter±1 µs
Trigger Latency5 µs ± 1 µs
Minimum Trigger Pulse Width15 ns
External Trigger Input LevelTTL compatible (Logic High > 2.0 V, Logic Low < 0.8 V)
External Trigger Output LevelTTL compatible (Logic High > 2.4 V, Logic Low < 0.4 V)
External Trigger Input Impedance10 kΩ (nominal)

Technology

Thermocouple

Applications

RF and Microwave Power Measurement