The Rohde & Schwarz FSW50 is a high-performance signal and spectrum analyzer designed to meet the demanding requirements of microwave and millimeter-wave applications, offering a frequency coverage from 2 Hz to 50 GHz with outstanding phase noise and dynamic range.

Specifications

General

ParameterValue
InterfacesLAN (LXI), GPIB (IEC 625-2/IEEE-488), USB 2.0, USB 3.0, DVI-D, DisplayPort
Display12.1-inch WXGA color touchscreen (1280 × 800 pixels)
Warm-Up Time20 min (10 min with Option FSW-B4)

Physical

ParameterValue
Power Supply100 V to 240 V AC (±10%), 50 Hz to 60 Hz / 400 Hz
Operating Temperature+5 °C to +45 °C
Dimensions (W×H×D)462 mm × 240 mm × 504 mm
Weight23.6 kg (configuration dependent)

Frequency Specifications

ParameterValue
Frequency Range2 Hz to 50 GHz
Frequency Resolution0.01 Hz
Frequency Span Range0 Hz (zero span), 10 Hz to 50 GHz
Internal Reference Frequency Aging Rate±1 × 10^-7 per year
Internal Reference Frequency Aging Rate (with Option FSW-B4)±3 × 10^-8 per year
Reference Temperature Stability±1 × 10^-9 (+5 °C to +45 °C, with Option FSW-B4)
Phase Noise at 10 kHz Offset (1 GHz Carrier)-140 dBc/Hz (typical)
Phase Noise at 100 kHz Offset (1 GHz Carrier)-143 dBc/Hz (typical)
Phase Noise at 1 MHz Offset (1 GHz Carrier)-150 dBc/Hz (typical)

Amplitude & Sensitivity Specifications

ParameterValue
RF Input Impedance50 Ω
RF Input Attenuator Range0 dB to 75 dB (in 5 dB steps)
RF Input Attenuator Step Size (with Option FSW-B25)1 dB
Maximum Safe Input Level (CW Power)+30 dBm (1 W)
Maximum Safe Input Level (DC Voltage)0 V

Connectivity & Interface Ports

ParameterValue
RF Input Connector Type2.4 mm female (ruggedized)
External Reference Input Frequency1 MHz to 20 MHz (in 1 Hz steps)
External Reference Output10 MHz (plus 100 MHz/640 MHz option)

Dynamic Range & Distortion

ParameterValue
Third-Order Intercept (TOI) at 2 GHz+25 dBm (typical)
1 dB Gain Compression Level (at 1 GHz)+15 dBm (nominal)

Sweep & Bandwidth

ParameterValue
Sweep Time Range (Zero Span)1 µs to 16000 s
Sweep Time Range (Span >= 10 Hz)1 µs to 16000 s
Resolution Bandwidth (RBW) Range1 Hz to 10 MHz (up to 80 MHz with options)
Video Bandwidth (VBW) Range1 Hz to 10 MHz
Standard Analysis Bandwidth80 MHz (configuration dependent)
Maximum Analysis Bandwidth8.3 GHz (with Option FSW-B8001)
RBW Selectivity (60 dB / 3 dB Shape Factor)< 4.6 (Gaussian shape)

Measurement Capabilities & Features

ParameterValue
DetectorsMax Peak, Min Peak, Auto Peak, Sample, RMS, Average, Quasi-Peak
Trace FunctionsClear/Write, Max Hold, Min Hold, Average, View, Blank
Number of Displayed Traces6
Standard Measurement SuitesChannel Power, ACPR, Occupied Bandwidth, Spectrum Emission Mask, CCDF

Physical & Environmental

ParameterValue
Display Resolution1280 × 800 pixels
Storage Temperature Range-40 °C to +70 °C
Operating Humidity Range95% RH (non-condensing)
Calibration Interval1 year

Receiver & Noise Performance

ParameterValue
DANL at 1 GHz (no preamplifier)-152 dBm (1 Hz) (typ. -155 dBm)
DANL at 10 GHz (no preamplifier)-150 dBm (1 Hz) (typ. -152 dBm)
DANL at 50 GHz (no preamplifier)-137 dBm (1 Hz) (typ. -141 dBm)
DANL at 50 GHz (with preamplifier Option FSW-B24)-156 dBm (1 Hz) (typ. -161 dBm)

Measurement Speed & Sweep

ParameterValue
Number of Sweep Points (Sweep)101 to 100001
Number of Sweep Points (Transient Recorder)2 to 400000000

Analysis Bandwidth

ParameterValue
Analysis Bandwidth Optionsup to 8.3 GHz (with Option FSW-B8000 / FSW-B8001)
Real-time Analysis Bandwidth Optionsup to 800 MHz (with Option FSW-B800R)

General Data

ParameterValue
Operating SystemWindows 10

Trigger

ParameterValue
Trigger SourcesFree Run, Video, External, IF Power, RF Power, I/Q Power, Timer

Connectivity

ParameterValue
Front-Panel USB Ports2 × USB 2.0 (Type A)
Rear-Panel USB Ports2 × USB 3.0 (Type A), 1 × USB 2.0 (Type B)
LAN Interface1 × RJ-45, 10/100/1000BASE-T
External Monitor Outputs1 × DVI-D, 1 × DisplayPort
Audio Output Port3.5 mm jack (front panel) for AM/FM demodulated audio

Inputs & Outputs

ParameterValue
Noise Source Control PortBNC, 28 V DC, max. 100 mA
External Mixer LO/IF PortOption FSW-B21 (SMA female, LO: 8.0 GHz to 14.5 GHz, IF: 311.25 MHz)
Standard IF Output Frequency311.25 MHz (BNC, 50 Ω)

Environmental & Safety

ParameterValue
Sinusoidal Vibration Compliance5 Hz to 55 Hz, 0.15 mm amplitude (meets EN 60068-2-6)
Random Vibration Compliance10 Hz to 300 Hz, 1.2 g RMS (meets EN 60068-2-64)
Shock Resistance Compliance40 g shock spectrum (meets MIL-STD-810G, Method 516.6)

Technology

Superheterodyne Spectrum Analyzer with Digital IF

Applications

Microwave and millimeter-wave testing, radar analysis, 5G and satellite communication systems