The Rohde & Schwarz NAS is a high-precision directional power meter designed for measuring transmitter power, reflected power, and Standing Wave Ratio (SWR) of transmitter systems. Equipped with special insertion units, it is highly suited for testing analog and digital mobile radio equipment, including GSM networks, and supports peak envelope power (PEP) and average power (AVG) measurements.

Specifications

General

ParameterValue
InterfacesRS-232-C (optional via NAS-B1 interface option)
DisplayTwo-line LC display with digital readouts (3.5-digit and 3-digit) and dual analog bar graphs

Physical

ParameterValue
Power Supply6 × 1.5 V AA dry cells, rechargeable NiCd battery pack, or external AC mains adapter
Operating Temperature0 to +50 °C
Dimensions (W×H×D)156 mm × 86 mm × 205 mm
Weight1.2 kg

Measurement Capabilities

ParameterValue
Frequency Range1 MHz to 2000 MHz (insertion unit dependent)
Power Measurement Range10 mW to 1200 W
SWR Measurement Range1.00 to 99.9
Return Loss Measurement Range0 to 40 dB
Measurement ParametersForward power, reflected power, standing wave ratio (SWR), reflection coefficient, return loss
Impedance50 Ω (determined by connected insertion unit)

User Interface and Display

ParameterValue
Display Resolution (Return Loss)0.1 dB
Display Measurement UnitsW, dBm, dB, SWR

Measurement Accuracy and Calibration

ParameterValue
SelftestIntegrated automatic self-test routine during power-on
AutorangingAutomatic measurement range selection

System Connectivity and Interfaces

ParameterValue
Optional Interface OptionRS-232-C Interface (Option NAS-B1)

Physical and Power Specifications

ParameterValue
Battery Operating Time (Alkaline)Approx. 70 hours
Battery Operating Time (NiCd)Approx. 18 hours
Power Saving FunctionAutomatic switch-off function to conserve battery power

Sensor Compatibility and Connection

ParameterValue
Compatible Insertion Unit NAS-Z1 Frequency Range1 MHz to 30 MHz
Compatible Insertion Unit NAS-Z1 Power RangeUp to 1200 W
Compatible Insertion Unit NAS-Z2 Frequency Range30 MHz to 120 MHz
Compatible Insertion Unit NAS-Z2 Power RangeUp to 120 W
Compatible Insertion Unit NAS-Z3 Frequency Range70 MHz to 1000 MHz
Compatible Insertion Unit NAS-Z3 Power RangeUp to 120 W
Compatible Insertion Unit NAS-Z5 Frequency Range70 MHz to 1000 MHz
Compatible Insertion Unit NAS-Z5 Power RangeUp to 1200 W
Compatible Insertion Unit NAS-Z7 Frequency Range800 MHz to 2000 MHz
Compatible Insertion Unit NAS-Z7 Power RangeUp to 100 W
Compatible Insertion Unit NAS-Z21 StandardGSM 900 applications
Compatible Insertion Unit NAS-Z22 StandardGSM 1800 (DCS 1800) and GSM 1900 (PCS 1900) applications

Environmental

ParameterValue
Storage Temperature Range-40 °C to +70 °C

Power & Battery

ParameterValue
Battery Type6 x 1.5 V AA (LR6) dry cells or 6 x 1.2 V NiCd accumulators
Auto-Off Delay2.5 min / 10 min / continuous (selectable)
Current Consumption (without backlighting)approx. 40 mA
Current Consumption (with backlighting)approx. 120 mA
External Power Supply Input7 V to 15 V DC
NiCd Battery ChargerIntegrated, active when external power supply is connected

Options

ParameterValue
IEC/IEEE-488 Bus Interface OptionNAS-B2
RS-232 Interface OptionNAS-B4

Compatible Insertion Units

ParameterValue
Compatible Insertion Unit NAS-Z12 Frequency Range100 MHz to 2000 MHz
Compatible Insertion Unit NAS-Z12 Power Range0.1 W to 120 W
Insertion Unit NAS-Z1 Insertion Loss< 0.05 dB
Insertion Unit NAS-Z1 SWR< 1.05
Insertion Unit NAS-Z1 Directivity≥ 35 dB
Insertion Unit NAS-Z2 Insertion Loss< 0.1 dB
Insertion Unit NAS-Z2 SWR< 1.06 (up to 500 MHz), < 1.12 (500 to 1000 MHz)
Insertion Unit NAS-Z2 Directivity≥ 30 dB (30 to 500 MHz), ≥ 26 dB (500 to 1000 MHz)

Display

ParameterValue
Digital Display Refresh Rate3 measurements/s
Analog Display Bargraph31-segment scale

Measurement Performance

ParameterValue
Meter Intrinsic Error±0.5% of reading ± 1 digit

Technology

RF Power Measurement

Applications

RF Transmitter Testing and Mobile Radio SWR/Power Diagnostics