The Rohde & Schwarz NRVS is a highly accurate, single-channel RF power meter designed for a wide range of power measurements from DC up to 40 GHz, depending on the connected sensor. Offering compatibility with intelligent NRV-Z and URV5-Z sensors, it features automated configuration, non-volatile memory for setups, and specialized calculations for pulse peak power.

Specifications

General

ParameterValue
InterfacesGPIB (IEEE-488 / IEC-625)
Display4.5-digit LC display

Physical

ParameterValue
Power Supply100 V / 120 V / 220 V / 240 V (±10%), 47 Hz to 440 Hz, 12 VA
Operating Temperature0 to +50 °C
Dimensions (W×H×D)219 mm × 103 mm × 240 mm
Weight2.5

Measurement Capabilities

ParameterValue
Frequency Range0 Hz (DC) to 40 GHz (sensor dependent)
Power Measurement Range-70 dBm to +45 dBm (sensor dependent)
Number of Channels1
Measurement ModesAverage power, calculated pulse peak power, Peak Envelope Power (PEP with NRV-Z31 / NRV-Z33)
Pulse Duty Factor Range0.0001 % to 100 %

Measurement Accuracy and Performance

ParameterValue
Display Resolution Steps3 selectable steps (0.1, 0.01, 0.001 dB)
Noise Suppression Filters13 digital filters (automatic or manual selection)
Zero AdjustmentAutomatic (initiated by key or GPIB command)

Reference Calibration Source

ParameterValue
Optional Calibration SourceNRVS-B1 (Sensor Check Source)
Calibration Source Frequency50 MHz
Calibration Source Power Level1.00 mW (0 dBm)
Calibration Source ConnectorN female

Sensor Compatibility

ParameterValue
Supported Sensor FamiliesR&S NRV-Z, URV5-Z
Sensor Calibration Data MemoryAutomatic readout of calibration data from sensor memory

Connectivity and Interfaces

ParameterValue
GPIB InterfaceIEEE-488.2 / IEC-625
Analog Recorder Output0 to 3 V
DC Frequency InputBNC, 0 to 10 V (for tracking frequency-response correction)

System and Control

ParameterValue
Save/Recall Memory Registers20 complete instrument setups
Operation ModeKeyboard operated (no softkeys or menus)

Physical and Environmental

ParameterValue
Storage Temperature Range-40 °C to +70 °C
Form Factor1/2 19" rack width, 2 HU

Power Supply

ParameterValue
Power Consumptionapprox. 12 VA
AC Input Voltage100 V / 120 V / 220 V / 240 V (±10%, selectable)
AC Input Frequency47 Hz to 440 Hz

Calibration Source

ParameterValue
Calibration Source SWR< 1.05 (with option NRVS-B1)

Performance

ParameterValue
Measurement Rate (Analog Display)15 measurements/s
Measurement Rate (Digital Display)5 measurements/s
Measurement Rate (GPIB)up to 20 measurements/s
Warm-up Time1 hour

Compliance

ParameterValue
Safety StandardEN 61010-1
RF Interference SuppressionEN 55011 Class B

Accuracy & Performance

ParameterValue
Instrumentation Uncertainty±0.5% (0.022 dB) at 23 °C ± 5 °C
Temperature Coefficient of Instrumentation< 0.1% per K (0.005 dB per K)

Inputs & Outputs

ParameterValue
Sensor ConnectorLEMO 12-pin (Series 2C) on front panel

Display & Readings

ParameterValue
Supported UnitsW, dBm, dB, dBµV, V, %

Environmental

ParameterValue
Maximum Operating Altitude4600 m
Maximum Storage Altitude12000 m
Humidity ResistanceUp to 95% relative humidity at +40 °C, non-condensing (compliant with IEC 68-2-30)

Compliance & Standards

ParameterValue
Electromagnetic Compatibility (EMC)Compliant with EN 55011 Class B and EN 50082-1
Vibration Resistance5 to 55 Hz (compliant with IEC 68-2-6)
Shock Resistance40g shock spectrum (compliant with IEC 68-2-27)

Options

ParameterValue
Rear Sensor Connector Option (NRVS-B5)Relocates the sensor connector from the front panel to the rear panel

Technology

RF and Microwave Power Meter

Applications

RF power measurement, peak envelope power calculation, and RF sensor calibration