The Rohde & Schwarz ZNB40 is a high-performance Vector Network Analyzer operating from 100 kHz to 40 GHz. Designed for demanding RF development and production environments, it features wide dynamic range, fast sweep speeds, outstanding temperature stability, and a modern 12.1" multi-touch touchscreen user interface.

Specifications

General

ParameterValue
InterfacesUSB (Host & Device), LAN (Gigabit Ethernet), GPIB (optional), External Monitor (DVI-D / DisplayPort depending on HW), Trigger I/O, Reference I/O
Display12.1" WXGA color TFT touchscreen, 1280 × 800 resolution

Physical

ParameterValue
Power Supply100 V to 240 V AC (±10%), 50 Hz to 60 Hz / 400 Hz, Max. 350 W (2-port) / 450 W (4-port)
Operating Temperature+5 °C to +40 °C
Dimensions (W×H×D)461 mm × 240 mm × 351 mm
Weight16.2 kg (2-port configuration)

Frequency & Stimulus

ParameterValue
Frequency range10 MHz to 40 GHz
Frequency resolution1 Hz
Frequency aging rate< 1 × 10^-6 per year (standard); < 1 × 10^-7 per year (with precision OCXO R&S ZNB-B4)
Number of test ports2 or 4 (hardware option dependent)
Test port impedance50 Ω
Power resolution0.01 dB

Receiver & Dynamic Range

ParameterValue
Maximum RF input level (damage level)+27 dBm
Maximum DC input voltage30 V
Trace noise magnitude< 0.005 dB RMS (at 10 kHz IF bandwidth)
Trace noise phase< 0.05° RMS (at 10 kHz IF bandwidth)

Measurement Speed & Settings

ParameterValue
Number of measurement points per sweep1 to 100,001
IF bandwidth range1 Hz to 10 MHz
Averaging factors1 to 1,000,000
Sweep typesLinear, logarithmic, segmented, power

Measurement Capabilities & Calibration

ParameterValue
Measured parametersS-parameters, wave quantities, ratios, impedance, admittance
Display formatsdB Mag, Phase, Group Delay, Smith Chart, Polar, SWR, Linear Mag, Real, Imaginary
Supported calibration typesTOSM (SOLT), TRL, LRL, TRM, TNA, UOSM
Electronic calibration (ECal) supportYes (compatible with R&S ZN-Zxx calibration units)
Time domain analysis capabilityYes (with option R&S ZNB-K2)
Frequency offset measurementsYes (with option R&S ZNB-K4)

Hardware Interfaces & Connections

ParameterValue
Test port connector type2.92 mm, male (ruggedized)
External reference input frequency10 MHz
External reference input level-3 dBm to +20 dBm (at 50 Ω)
Internal reference output frequency10 MHz
Internal reference output level+9 dBm ±4 dB (into 50 Ω)
External trigger input levelTTL compatible (BNC female)
GPIB interfaceOptional (R&S ZNB-B10)
LAN interfaceRJ-45 (10/100/1000 Base-T)
USB ports4 × USB 2.0 (front), 4 × USB (rear, hardware version dependent)
External monitor outputDVI-D, DisplayPort (hardware version dependent)

System & Storage

ParameterValue
Display type12.1" WXGA color TFT touchscreen
Display resolution1280 × 800 pixels
Internal data storageSSD (Solid State Drive)
Data storage formats.s1p, .s2p, .s3p, .s4p, .csv, .dat, .png, .bmp
Remote control command languageSCPI
Internal operating systemWindows (hardware version dependent)

Physical, Environmental & Power

ParameterValue
Storage temperature range-40 °C to +70 °C
Operating humidity rangeNon-condensing, compliant with EN 60068-2-30
Line voltage frequency50 Hz to 60 Hz / 400 Hz
Line voltage range100 V to 240 V AC (±10%)
Maximum power consumption450 W (2-port), 550 W (4-port)

Display & User Interface

ParameterValue
Display size12.1 in (30.7 cm)

Source Characteristics

ParameterValue
Harmonics< -30 dBc (10 MHz to 100 MHz); < -35 dBc (100 MHz to 40 GHz) at -10 dBm

Receiver Characteristics

ParameterValue
Receiver 1 dB compression point> +10 dBm (typ. +13 dBm) at test port

Measurement Capabilities

ParameterValue
Max. number of channels100
Max. number of traces per channel100
Max. number of diagrams100

Hardware Options

ParameterValue
Integrated bias tees (Option R&S ZNB-B1)Max. 30 V, max. 250 mA (integrated as standard)
Precision frequency reference (Option R&S ZNB-B4)Aging < 1 × 10^-9 per day, temperature drift < 1 × 10^-8 (option-dependent)
Direct generator/receiver access (Option R&S ZNB-B16)Provides direct access to generator outputs and receiver inputs (option-dependent)

Rear Panel Interfaces

ParameterValue
User port (Handler I/O)25-pin D-Sub female connector
External trigger outputBNC female, TTL-compatible

Integrated Bias Tees

ParameterValue
Bias tee maximum voltage (Option R&S ZNB-B1)30 V (integrated as standard)
Bias tee maximum current (Option R&S ZNB-B1)250 mA (integrated as standard)

Compliance

ParameterValue
LXI complianceLXI Version 1.4 Class C

Calibration

ParameterValue
Automatic calibration unit supportSupported (R&S ZN-Zxx series)

Display

ParameterValue
Multi-touch touchscreenSupported

Options

ParameterValue
Removable hard disk optionSupported (Option R&S ZNB-B19)

Software Options

ParameterValue
Distance-to-fault measurementsSupported (Option R&S ZNB-K3)
Frequency conversion measurementsSupported (Option R&S ZNB-K4)
Intermodulation measurementsSupported (Option R&S ZNB-K14)
Noise figure measurementsSupported (Option R&S ZNB-K30)
In-situ calibrationSupported (Option R&S ZNB-K210)

Technology

Vector Network Analysis

Applications

RF and Microwave Component Characterization

Looking for the Rohde & Schwarz ZNB40?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us