The Rohde & Schwarz ZNB8 is a high-performance vector network analyzer designed for demanding RF component development, production, and service applications. Operating from 9 kHz to 8.5 GHz, it features exceptional measurement speed, high dynamic range, and outstanding temperature stability.

Specifications

General

ParameterValue
InterfacesLAN, USB, GPIB (optional)
Display12.1" high-resolution touchscreen

Physical

ParameterValue
Operating Temperature+5 °C to +40 °C
Dimensions (W×H×D)461 mm × 240 mm × 351 mm
Weight13.7 kg

RF Performance & Port Characteristics

ParameterValue
Frequency Range Min9 kHz
Frequency Range Max8.5 GHz
Frequency Resolution1 Hz
Test Port Connector TypeN female
Test Port Impedance50 Ω
Test Port Input Match (9 kHz to 20 kHz)>10 dB
Test Port Input Match (20 kHz to 8.5 GHz)>20 dB
Output Power Range (2.5 GHz to 7.5 GHz, Standard)–55 dBm to +10 dBm
Typical Output Power (2.5 GHz to 7.5 GHz, Standard)+13 dBm
Output Power Range (7.5 GHz to 8.5 GHz, Standard)–55 dBm to +8 dBm
Typical Output Power (7.5 GHz to 8.5 GHz, Standard)+12 dBm
Output Power Range (2.5 GHz to 7.5 GHz, with Option B22)–85 dBm to +10 dBm
Typical Output Power (2.5 GHz to 7.5 GHz, with Option B22)+13 dBm
Output Power Range (7.5 GHz to 8.5 GHz, with Option B22)–85 dBm to +8 dBm
Typical Output Power (7.5 GHz to 8.5 GHz, with Option B22)+12 dBm

Receiver Characteristics & Dynamic Range

ParameterValue
System Dynamic Range (at 10 Hz IF Bandwidth)up to 140 dB
Dynamic Range (9 kHz to 100 kHz)>100 dB
Typical Dynamic Range (9 kHz to 100 kHz)122 dB
Dynamic Range (100 kHz to 50 MHz)>120 dB
Typical Dynamic Range (100 kHz to 50 MHz)138 dB
Dynamic Range (50 MHz to 4 GHz)>130 dB
Typical Dynamic Range (50 MHz to 4 GHz)140 dB
Dynamic Range (4 GHz to 7 GHz)>125 dB
Typical Dynamic Range (4 GHz to 7 GHz)138 dB
Test Port Damage Level+27 dBm
Test Port Damage DC Voltage30 V
IF Bandwidth Range (Standard)1 Hz to 1 MHz
IF Bandwidth Range (with Increased Bandwidth option)1 Hz to 10 MHz

Measurement Capabilities & Formats

ParameterValue
Number of Sweep Points per Trace2 to 100001
Measured ParametersMixed-mode S-parameters
Time-Domain Analysis (DTF)Supported

Calibration & Accuracy Enhancement

ParameterValue
Calibration Methods SupportedManual and automatic

System & Connectivity Interfaces

ParameterValue
Touchscreen CapabilityYes
Display Resolution1280 × 800 pixels (WXGA)

Software & Functional Options

ParameterValue
Operating SystemWindows

Reference Frequency

ParameterValue
Frequency Aging per Year (Standard)±1 × 10⁻⁶
Frequency Aging per Year (with Option B4)±1 × 10⁻⁷
Frequency Temperature Stability (Standard)±1 × 10⁻⁶ (+5 °C to +40 °C)
Frequency Temperature Stability (with Option B4)±1 × 10⁻⁸ (+5 °C to +40 °C)

Interfaces

ParameterValue
External Monitor PortDVI-D
Number of USB Ports (Front)4 (USB 2.0)

Technology

Vector Network Analysis

Applications

RF component testing, mobile radio, and electronic goods manufacturing