The Tektronix 2445 is a high-performance, 150 MHz, four-channel analog oscilloscope designed for general-purpose, laboratory, and field service applications. Known for its advanced CRT readout, cursor measurement system, and rugged construction, the 2445 provides detailed waveform analysis with exceptional precision.

Specifications

General

ParameterValue
InterfacesGPIB (optional, Option 10)
DisplayCRT (Green Phosphor, P31, with CRT Readout)

Physical

ParameterValue
Power Supply90 V to 250 V AC, 48 Hz to 440 Hz, 120 W max
Operating Temperature-15 °C to +55 °C
Weight9.3 kg

Vertical System

ParameterValue
Bandwidth (-3 dB)150 MHz
Rise Time2.33 ns
Number of Channels4
Deflection Factor Range (CH 1 & CH 2)2 mV/div to 5 V/div
Deflection Factor Range (CH 3 & CH 4)0.1 V/div and 0.5 V/div
Vertical Accuracy±2%
Input Impedance (CH 1 & CH 2)1 MΩ || 15 pF (selectable to 50 Ω)
Input Impedance (CH 3 & CH 4)1 MΩ || 15 pF
Maximum Input Voltage (1 MΩ)400 V (DC + peak AC)
Maximum Input Voltage (50 Ω)5 V RMS
Input Coupling (CH 1 & CH 2)AC, DC, GND
Input Coupling (CH 3 & CH 4)DC only
Vertical Display ModesCH 1, CH 2, ADD, ALT, CHOP, CH 3, CH 4
Invert Channel CapabilityCH 2
Channel Isolation (Crosstalk)100:1 or greater at 100 MHz
Common Mode Rejection Ratio (CMRR)At least 20:1 at 50 MHz
Bandwidth Limit (-3 dB)20 MHz
Vertical Chop Frequency1 MHz
AC Coupling Low-Frequency Limit (-3 dB)10 Hz or less (1X probe), 1 Hz or less (10X probe)
Probe Scale Factor SensingAutomatic detection of 1X, 10X, or 100X coded probes
Input Capacitance (CH 1 & CH 2)15 pF ± 2 pF
Input Capacitance (CH 3 & CH 4)15 pF ± 3 pF

Horizontal System

ParameterValue
A Sweep Rate Range20 ns/div to 0.5 s/div
B Sweep Rate Range20 ns/div to 50 ms/div
Sweep Rate Accuracy±0.7%
Sweep Magnifierx10
Horizontal Display ModesA, A INTEN, ALT, B (Delayed)

Trigger System

ParameterValue
Trigger ModesAuto Level, Auto, Normal, Single Sequence
Trigger SourcesCH 1, CH 2, CH 3, CH 4, Line, Vertical
Trigger Coupling ModesAC, DC, HF Reject, LF Reject, Noise Reject
Trigger Level Range±18 times the VOLTS/DIV setting (CH 1 & CH 2)
Trigger Sensitivity (CH 1 & CH 2, DC Coupled)0.35 div from DC to 30 MHz, 1.5 div at 150 MHz

X-Y Operation

ParameterValue
X-Axis Input SourceCH 1
Y-Axis Input SourceCH 2
X-Axis BandwidthDC to 3 MHz
X-Y Phase DifferenceWithin 3° from DC to 1 MHz

CRT & Display

ParameterValue
CRT Size8 cm x 10 cm
Phosphor TypeP31 (standard)
Graticule Illumination ControlAdjustable
Trace Rotation AdjustmentFront panel screwdriver-adjustable

Cursor & Readout System

ParameterValue
CRT Readout CapabilityVolts, Time, Frequency, Phase, Ratio, Trigger Level
Cursor Measurement FunctionsΔV, Δt, 1/Δt (frequency)

Auxiliary Outputs & Calibrator

ParameterValue
Calibrator Output Voltage (Amplitude)0.5 V square wave (±1%) into high impedance
Calibrator Output Frequency1 kHz (±1%)
CH 2 Vertical Signal Output Voltage20 mV/div into 50 Ω
A Gate Output VoltageTTL compatible (active high)
B Gate Output VoltageTTL compatible (active high)

Power Supply

ParameterValue
AC Input Voltage Range90 V to 250 V AC
AC Input Frequency Range48 Hz to 440 Hz
Power Consumption120 W maximum

Physical Specifications

ParameterValue
Width338 mm
Height137 mm
Depth445 mm
Net Weight9.3 kg
Shipping WeightApprox. 12.7 kg
Cooling System TypeForced-air cooling (fan)

Environmental & Reliability

ParameterValue
Operating Temperature Range-15 °C to +55 °C
Non-Operating Temperature Range-62 °C to +85 °C
Operating Altitude LimitUp to 4,600 m (15,000 ft)
Non-Operating Altitude LimitUp to 15,200 m (50,000 ft)
Vibration ResistanceMeets MIL-T-28800C, Class 3
Shock ResistanceMeets MIL-T-28800C, Class 3

Display

ParameterValue
CRT Accelerating Voltage16 kV
Beam FinderCompresses the display to within the graticule area and intensifies the trace
Graticule TypeInternal, parallax-free

Options

ParameterValue
Option 01 (Digital Multimeter)4.5-digit autoranging DMM, measures DCV, ACV, Ohms, and temperature
Option 05 (TV Triggering)Compatible with SECAM, NTSC, and PAL standards; provides TV Field 1, Field 2, and Line Sync
Option 06 (Counter/Timer/Trigger - CTT)Provides 9-digit frequency counter, period, and time interval measurements up to 150 MHz
Option 09 (Word Recognizer)Adds a 17-bit word recognizer probe (16 data + 1 clock) to the Counter/Timer/Trigger
Option 10 (GPIB Interface)Provides full remote programmability conforming to IEEE-488.1
Option 11 (Probe Power)Provides power for active probes (+5 V, -5 V, +15 V, and -15 V)
Option 1R (Rackmount)Configures the instrument for installation in a standard 19-inch rack

Compliance

ParameterValue
Safety StandardsCertified to UL 1244, CSA C22.2 No. 231
Military SpecificationsMeets MIL-T-28800C, Class 5 environmental limits

Z-Axis System

ParameterValue
Z-Axis Input Impedance9.1 kΩ

Calibrator

ParameterValue
Calibrator Loop Current10 mA (into shorting loop)

Technology

Analog

Applications

Electronic testing and measurement