The Tektronix DPO3014 is a 100 MHz, 4-channel digital phosphor oscilloscope from the DPO3000 Series. Featuring a standard 5 Mpoint record length and up to 2.5 GS/s sampling rate across all channels, it delivers exceptional signal fidelity and capture depth. The integrated Wave Inspector® controls provide efficient search and navigation through long records, while the 9-inch WVGA widescreen display and TekVPI® probe interface make it highly versatile for debug and analysis tasks.

Specifications

General

ParameterValue
InterfacesUSB 2.0 Host (x2), USB 2.0 Device, LAN (10/100 Base-T), SVGA Video Out
Display9.0-inch (229 mm) WVGA widescreen TFT color LCD, 800 x 480 pixels
Calibration Interval1 year

Physical

ParameterValue
Power Supply100 V to 240 V ±10% (50 Hz to 60 Hz), 115 V ±10% (400 Hz), 120 W max
Operating Temperature0 °C to +50 °C
Dimensions (W×H×D)416.6 mm x 203.2 mm x 137.2 mm
Weight4.2 kg

Vertical System

ParameterValue
Analog Bandwidth (-3 dB)100 MHz
Calculated Rise Time (10% to 90%)3.5 ns
Number of Channels4
Input Impedance1 MΩ ± 1% in parallel with 13 pF ± 2 pF, 50 Ω ± 1%
Input CouplingAC, DC
Vertical Resolution8 bits
Maximum Input Voltage (1 MΩ)300 Vrms CAT II with peaks ≤ ±425 V
Maximum Input Voltage (50 Ω)5 Vrms with peaks ≤ ±20 V
Vertical Sensitivity Range (1 MΩ)1 mV/div to 10 V/div
Vertical Sensitivity Range (50 Ω)1 mV/div to 1 V/div
Hardware Bandwidth Limits20 MHz
Channel-to-Channel Isolation≥100:1 at ≤100 MHz
Vertical Offset Range±1 V to ±100 V (scale and impedance dependent)

Horizontal System

ParameterValue
Maximum Real-Time Sampling Rate2.5 GS/s on all channels
Maximum Record Length5 Mpoints on all channels
Time Base Range4 ns/div to 1000 s/div
Time Base Accuracy±10 ppm
Channel-to-Channel Deskew Range±75 ns
Timebase Delay Time Range-10 divisions to 5000 s

Trigger System

ParameterValue
Trigger ModesAuto, Normal, Single
Trigger CouplingDC, AC, HF Reject, LF Reject, Noise Reject
Trigger Holdoff Range20 ns to 8 s
Trigger TypesEdge, Sequence, Pulse Width, Runt, Logic, Setup/Hold, Rise/Fall Time, Video, I2C/SPI/CAN/LIN/RS-232/422/485/UART (optional)
Trigger Sensitivity (Internal, DC Coupled)0.4 div from DC to 50 MHz, 1 div at 100 MHz
Trigger Level RangeAny channel: ±8 divisions from center of screen
Internal Trigger Level Range±8 divisions from screen center
Trigger Level Accuracy±0.2 div (typical, for signals with rise/fall times ≥ 20 ns)

Acquisition System

ParameterValue
Maximum Waveform Capture Rate50,000 wfm/s
Acquisition ModesSample, Peak Detect, Averaging, Envelope, Hi-Res, Roll
Averaging Factor Range2 to 512
Minimum Peak Detect Pulse Width4.0 ns
Hi-Res Mode Vertical ResolutionUp to 11 bits

Waveform Measurements and Analysis

ParameterValue
Automatic Measurements29 automatic measurements
Waveform MathArithmetic (Add, Subtract, Multiply), FFT
FFT Window TypesRectangular, Hamming, Hanning, Blackman-Harris

Display and User Interface

ParameterValue
Display Size9.0 in. (229 mm)
Display Resolution800 x 480 pixels (WVGA)

Connectivity and Ports

ParameterValue
Probe InterfaceTekVPI Probe Interface
USB Host Ports2 (1 front, 1 rear)
USB Device Ports1 (rear, USBTMC)
LAN Port (Ethernet)1 (RJ-45, 10/100 Base-T)
Video Output Port1 (DB-15 female, SVGA)
Auxiliary OutputBNC (Trigger Out)

Environmental Specifications

ParameterValue
Storage Temperature Range-40 °C to +71 °C
Operating Humidity RangeUp to 60% RH at +40 °C to +50 °C, Up to 90% RH at or below +40 °C
Operating AltitudeUp to 3,000 m

Physical Characteristics

ParameterValue
Form FactorBenchtop
Security Lock SlotKensington-style lock slot
Rackmount OptionRMD3000

Wave Inspector® Navigation and Search

ParameterValue
Wave Inspector® NavigationDedicated front-panel controls for zoom/pan, play/pause, search, and user marks

Serial Bus Triggering and Analysis

ParameterValue
Supported Protocols (with Optional Modules)I2C, SPI, CAN, LIN, RS-232/422/485/UART, I2S/LJ/RJ/TDM

Interfaces & Connectivity

ParameterValue
GPIB PortOptional via TEK-USB-488 adapter
LXI ComplianceLXI Class C

Power Specifications

ParameterValue
Power Consumption (Max)120 W

Standards & Compliance

ParameterValue
Safety StandardsUL61010-1, CAN/CSA-C22.2 No. 61010-1, EN 61010-1

Storage

ParameterValue
Reference Waveform Storage4 internal waveforms
Saved Setups Storage10 internal setups

Interfaces

ParameterValue
Ethernet Port Speed10/100 Mbps (10/100Base-T)
GPIB InterfaceOptional via TEK-USB-488 Adapter
Video Output ConnectionDB-15 female (SVGA)

Auxiliary Outputs

ParameterValue
Probe Compensator Output1 kHz, 0 to 2.5 V
Auxiliary Output SignalTrigger Out (TTL-compatible)

Physical Specifications

ParameterValue
Security LockRear-panel Kensington-style security slot

Compliance & Safety

ParameterValue
Electromagnetic Compatibility (EMC)EC Council Directive 2004/108/EC, EN61326-1:2006 Class A; AS/NZS 2064

Software & Connectivity

ParameterValue
Included SoftwareOpenChoice® Desktop, NI LabVIEW SignalExpress™ Tektronix Edition
Driver SupportIVI-COM, IVI-C, and LabVIEW drivers

Wave Inspector® Navigation

ParameterValue
Automated Search CriteriaEdge, pulse width, runt, logic, setup/hold, rise/fall time, packet content

Technology

Digital Phosphor Oscilloscope (DPO)

Applications

Electronic Design and Debug, Education and Training, Manufacturing Test and Quality Control