The Tektronix DPO70404 is a high-performance Digital Phosphor Oscilloscope (DPO) belonging to the DPO70000 Series. Designed for verifying, characterizing, debugging, and testing sophisticated electronic and high-speed serial designs, it offers a 4 GHz analog bandwidth and a real-time sample rate of 25 GS/s on all four channels simultaneously. It features Pinpoint triggering, a highly intuitive user interface with a large touch screen, and advanced software compatibility.

Specifications

General

ParameterValue
InterfacesUSB, LAN (10/100/1000 Base-T), GPIB, RS-232, DVI, VGA
Display12.1 in. TFT-LCD color touch screen, 1024 × 768 (XGA)
Calibration Interval1 year (recommended)
LXI ComplianceLXI Class C Version 1.3

Physical

ParameterValue
Power Supply100 to 240 V AC ±10%, 47 to 63 Hz (and 400 Hz at 115 V), <1200 VA
Operating Temperature5 °C to +45 °C
Dimensions (W×H×D)451 mm × 298 mm × 489 mm
Weight20 kg

Vertical System

ParameterValue
Input Channels4
Analog Bandwidth (-3 dB)4 GHz
Rise Time (10% to 90%)98 ps (typical)
Rise Time (20% to 80%)65 ps (typical)
Input Impedance50 Ω ±1.5%
Input CouplingDC, GND
Input Sensitivity Range10 mV/div to 1 V/div
Vertical Resolution8-bit (up to 11-bit with averaging)
DC Gain Accuracy±2%
Offset Range±1 V (10 to 99.5 mV/div), ±5 V (100 mV to 1 V/div)
Maximum Input Voltage5 V RMS, with peaks ≤ ±24 V
Channel-to-Channel Isolation≥ 120:1 at ≤ 100 MHz, ≥ 80:1 at > 100 MHz to 2.5 GHz, ≥ 50:1 at > 2.5 GHz to 4 GHz
Low-frequency Limit (AC Coupling)N/A (DC coupled only)

Horizontal System

ParameterValue
Maximum Real-Time Sample Rate25 GS/s (on all 4 channels)
Maximum Record Length (Standard)10 MS per channel (40 MS max)
Maximum Record Length (Optional Upgrade)up to 200 MS (option dependent)
Timebase Range20 ps/div to 1000 s/div
Channel-to-Channel Deskew Range±75 ns
Timebase Delay Range-5.0 ks to 250 s
Timebase Accuracy±1.5 ppm initial accuracy, with aging <1 ppm/year
Equivalent-Time Sample RateUp to 2.5 TS/s

Acquisition System

ParameterValue
Acquisition ModesSample, Peak Detect, Hi Res, Envelope, Average, Waveform Database, FastAcq
Maximum Waveform Capture Rate (FastAcq)>300,000 wfms/s (per channel)
FastFrame™ Segmented MemoryStandard, up to 1,000,000 frames (configuration and record length dependent)

Trigger System

ParameterValue
Trigger System TypePinpoint Triggering
Trigger Combinations>1400
Trigger ModesAuto, Normal, Single
Trigger CouplingDC, AC, Low Frequency Reject, High Frequency Reject, Noise Reject
Trigger Jitter (Typical)1 ps RMS
Serial Pattern Triggering Speedup to 3.125 Gb/s
Trigger Level Range (Internal)±12 divisions from center of screen
Trigger Holdoff Range250 ns to 12 s

Waveform Measurements and Analysis

ParameterValue
Automatic Measurements53 automatic measurements
Waveform HistogramsVertical and Horizontal
Algebraic Math FunctionsAdd, Subtract, Multiply, Divide
Advanced MathFFT, user-defined equations

Display System

ParameterValue
Display TypeLiquid crystal active-matrix color TFT
Display Size12.1 in.
Display Resolution1024 × 768 pixels
FormatYT, XY, XYZ

Computer System & Storage

ParameterValue
Operating SystemMicrosoft Windows XP
Storage Drive Capacity≥80 GB removable hard disk drive
Optical Drive TypeCD-RW/DVD drive

Connectivity & Interfaces

ParameterValue
Analog Input ConnectorsTekConnect
USB Ports5 USB 2.0 ports (1 front, 4 rear)
LAN Port1 RJ-45 (10/100/1000 Base-T)
GPIB InterfaceIEEE 488.2
External Video OutputDVI, VGA
External Reference Input10 MHz (BNC)
Internal Reference Output10 MHz (BNC)
Auxiliary Trigger OutputYes (BNC)

Power and Environmental

ParameterValue
Operating Humidity8% to 80% relative humidity, non-condensing
Operating Altitudeup to 3,000 m (9,843 ft.)
Non-Operating Altitudeup to 12,190 m (40,000 ft.)

Physical Characteristics

ParameterValue
Height298 mm
Width451 mm
Depth489 mm
Rackmount Configuration Size7U

Compliance & Certifications

ParameterValue
Safety Standards ComplianceUL61010-1, CSA22.2 No. 61010-1, EN61010-1
Electromagnetic Compatibility (EMC)EN61326, FCC Part 15 Class A

Software Options & Application Packages

ParameterValue
Jitter and Eye Diagram AnalysisOption DJA / DPOJET
Serial Data Compliance TestingYes (PCI Express, SATA, FB-DIMM, SAS, Fibre Channel, Ethernet, HDMI, DVI, USB 2.0)
DDR Analysis SoftwareOption DDRA (supports DDR1, LP-DDR1, DDR2, DDR3, GDDR3)
Limit TestingYes
Custom Interface SoftwareMyScope Custom Windows, OpenChoice Software

Power Specifications

ParameterValue
Input Voltage Range100 to 240 V AC ±10% (50 to 60 Hz), 115 V AC ±10% (400 Hz)

Inputs & Outputs

ParameterValue
Probe InterfaceTekConnect® interface, supporting active, differential, and optical probes
TekLink Port1 rear-panel proprietary TekLink connector

Environmental

ParameterValue
Cooling Clearance Requirements76 mm (3 in.) minimum on Left, Right, and Rear
Operating Shock15 g, 11 ms, half-sine

Power

ParameterValue
Power Line Frequency Range47 Hz to 63 Hz (90 V to 264 V) or 360 Hz to 440 Hz (100 V to 132 V)

Technology

Digital Phosphor

Applications

Signal integrity verification, jitter characterization, high-speed serial design debugging and compliance testing