The Tektronix MSO71254C is a high-performance 12.5 GHz Mixed Signal Oscilloscope belonging to the MSO70000C Series. Designed for deep system-level debug of high-speed digital designs, it features 4 analog channels and 16 logic channels. It delivers real-time sample rates of up to 100 GS/s on 2 channels (50 GS/s on all 4 channels) and a 12.5 GS/s sample rate with 80 ps timing resolution on digital channels. The instrument is equipped with Pinpoint triggering, a hardware-based 6.25 Gb/s serial pattern trigger, iCapture analog multiplexing of digital lines, and comprehensive serial decoding options.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | USB 2.0 Host, USB Device, LAN (10/100/1000 Base-T Ethernet), GPIB, LXI Class C, DVI-I, TekConnect |
| Display | 12.1-inch liquid-crystal active-matrix color TFT-LCD touchscreen (1024 × 768) |
| LXI Compliance | LXI Class C Version 1.3 |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 V AC ±10% (50/60 Hz), 115 V AC ±10% (400 Hz), <980 VA max |
| Operating Temperature | 5 to 45 °C |
| Dimensions (W×H×D) | 451 × 298 × 489 mm |
| Weight | 24 |
Analog Vertical System
| Parameter | Value |
|---|---|
| Analog Input Channels | 4 |
| Analog Bandwidth | 12.5 GHz (User-selectable DSP enhance) |
| Hardware Analog Bandwidth (-3 dB) | 12.5 GHz (typical) |
| Calculated Rise Time (10% to 90%) | 32 ps (typical) |
| Vertical Resolution | 8 bit (11 bit with averaging) |
| Max Input Voltage (50 Ohm) | <5.0 V_RMS for >=100 mV/div; 1.5 V_RMS for <100 mV/div |
Digital Vertical System
| Parameter | Value |
|---|---|
| Digital Channels | 16 |
| Vertical Resolution (Digital) | 1 bit |
| Logic Threshold Range (P6780) | -2 to +4.5 V |
| Logic Threshold Range (P6717) | -1.5 to +4.0 V |
| Maximum Input Voltage (Digital Probe) | ±15 V (nondestruct) |
| Digital Bandwidth (P6780) | 2.5 GHz |
| Digital Bandwidth (P6717) | 350 MHz |
Horizontal System
| Parameter | Value |
|---|---|
| Time Base Range | 10 ps/div to 1000 s/div |
| Time Resolution (ET/IT mode) | 100 fs |
| Channel-to-Channel Deskew Range | ±75 ns |
| Jitter Noise Floor | 270 fs (typical, with BW bandwidth enhance enabled) |
| Time Base Delay Range | -5.0 ks to 1.0 ks |
| Timebase Aging | ±1 ppm/year |
| Timebase Accuracy | ±1.5 ppm (initial) |
Acquisition System (Analog)
| Parameter | Value |
|---|---|
| Max Real-time Sample Rate (2 Channels) | 100 GS/s |
| Max Real-time Sample Rate (4 Channels) | 50 GS/s |
| Max Equivalent Time / Interpolated Time Sample Rate | 10 TS/s |
| Maximum Record Length (Standard Configuration) | 31.25 M on all 4 channels |
| Maximum Duration at Highest Real-time Resolution (Standard) | 0.625 ms |
Acquisition System (Digital)
| Parameter | Value |
|---|---|
| Max Digital Sample Rate (All Channels) | 12.5 GS/s |
| Digital Timing Resolution | 80 ps |
| Maximum Record Length (Digital, with option) | 250 M on all channels |
| Minimum Detectable Pulse Width | 300 ps |
| Maximum Number of Buses | 16 |
| Number of Channels per Bus | Up to 24 (16 logic, 4 analog, 4 math) |
Acquisition Modes
| Parameter | Value |
|---|---|
| Averaging Mode | From 2 to 10,000 waveforms |
| Envelope Mode | From 1 to 2x10^9 waveforms |
| FastAcq Waveform Capture Rate | >300,000 wfms/s on all 4 channels simultaneously |
| FastFrame Segmented Memory | Maximum trigger rate >310,000 waveforms per second |
| High Resolution (Hi-Res) Mode | Real-time boxcar averaging reduces random noise and increases resolution |
| Peak Detect Mode | Not supported |
| Roll Mode | Not supported |
| Sample Mode | Acquires and displays sampled values |
| Waveform Database Mode | Accumulates three-dimensional array of amplitude, time, and counts |
Trigger System
| Parameter | Value |
|---|---|
| Digital A then Analog B Triggering | Standard |
| A Event and Delayed B Event Trigger Types | Edge, Glitch, Runt, Width, Transition Time, Timeout, Pattern, State, Setup/Hold, Window |
| Main Trigger Modes | Auto, Normal, Single |
| Trigger Sequences | Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State, Reset by Transition |
| Serial Pattern Triggering | Up to 6.25 Gb/s (hardware-based, NRZ or 8b/10b patterns) |
| Embedded Serial Bus Support | I2C, SPI Decoding and Triggering |
| Trigger Jitter | <100 fs RMS (typical) |
| Trigger Level Range (Internal) | ±120% of full scale |
Waveform Analysis & Features
| Parameter | Value |
|---|---|
| iCapture Multiplexer Technology | Route any 4 of the 16 logic channels to the analog acquisition system |
| Probing System Support | P7500 TriMode Probing System (4 GHz to 20 GHz) |
Environmental & Compliance
| Parameter | Value |
|---|---|
| Operating Altitude | Up to 3,000 m (9,843 ft.) |
Vertical System - Analog
| Parameter | Value |
|---|---|
| Input Coupling (Analog) | DC |
| DC Gain Accuracy | ±2% |
| Vertical Sensitivity Range | 10 mV/div to 1 V/div |
Vertical System - Digital
| Parameter | Value |
|---|---|
| Digital Input Impedance (with P6717 Probe) | 20 kΩ || 3 pF |
| Digital Input Impedance (with P6780 Probe) | 20 kΩ || 0.8 pF |
| Digital Threshold Accuracy | ±(100 mV + 3% of threshold setting) |
| Digital Threshold Resolution | 5 mV |
| Minimum Digital Voltage Swing | 300 mVp-p |
Computer System
| Parameter | Value |
|---|---|
| Operating System | Windows 7 Ultimate (64-bit) |
Environmental
| Parameter | Value |
|---|---|
| Non-Operating Temperature | -20 °C to +60 °C |
| Operating Humidity | 8% to 80% RH (at or below +32 °C) |
| Non-Operating Humidity | 5% to 95% RH (at or below +40 °C) |
| Non-Operating Altitude | 12,190 m (40,000 ft) |
Vertical System (Analog)
| Parameter | Value |
|---|---|
| Rise Time (20% to 80%) | 22 ps |
| Analog Input Offset Range | ±0.5 V (10 mV/div to 99.5 mV/div), ±5.0 V (100 mV/div to 500 mV/div) |
| Position Range | ±5 divisions |
Waveform Math & Analysis
| Parameter | Value |
|---|---|
| Number of Math Waveforms | 4 |
| Math Functions | Arithmetic, algebraic, trigonometric, exponential, logarithmic, integration, differentiation, FFT |
| FFT Window Types | Rectangular, Hamming, Hanning, Kaiser-Bessel, Blackman-Harris, Gaussian, Flattop2, TekExponential |
| Automatic Measurements | 53 automatic measurements |
Connectivity & I/O
| Parameter | Value |
|---|---|
| USB Ports | 1x USB 2.0 host (front), 4x USB 2.0 host (rear), 1x USB device (rear) |
| Video Output Ports | 1x DVI-I, 1x VGA |
Horizontal System / Timebase
| Parameter | Value |
|---|---|
| External Reference Input Frequency | 10 MHz |
| External Reference Input Impedance | 50 Ohm |
Physical & Environmental
| Parameter | Value |
|---|---|
| Cooling Clearance Requirements | 76 mm (3 in.) on left, right, and rear |
| Operating Shock Resistance | 15 g, 11 ms half-sine |
| Non-Operating Shock Resistance | 30 g, 11 ms half-sine |
| Operating Vibration Resistance | 0.31 GRMS (5 Hz to 500 Hz) |
| Non-Operating Vibration Resistance | 2.46 GRMS (5 Hz to 500 Hz) |
Technology
Mixed Signal Oscilloscope
Applications
High-speed digital design, DDR memory verification, serial standards testing, digital RF, ASIC/FPGA debugging
Looking for the Tektronix MSO71254C?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.
Selling this kind of equipment? Sell with us