The Tektronix TDS2022 is a 200 MHz, 2-channel digital storage oscilloscope featuring a color TFT LCD, up to 2.0 GS/s real-time sample rate on all channels, and optional communication modules for PC connectivity and data storage.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | Expansion slot for optional modules (TDS2CMA: RS-232/Centronics, TDS2MEM: RS-232/Centronics/CompactFlash, TDSPACK: RS-232/Centronics/GPIB) |
| Display | 5.7 in (145 mm) Color Active Matrix TFT LCD, 320 × 240 pixels |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 to 240 VAC (±10%), 47 to 440 Hz (at 115 V), 47 to 63 Hz (at 230 V), <30 W power consumption |
| Operating Temperature | 0 to +50 °C |
| Dimensions (W×H×D) | 323.6 mm × 151.4 mm × 124.5 mm |
| Weight | 2.0 kg |
Vertical System
| Parameter | Value |
|---|---|
| Analog Bandwidth (-3 dB) | 200 MHz |
| Calculated Rise Time (10% to 90%) | 1.8 ns |
| Number of Input Channels | 2 |
| Input Impedance | 1 MΩ ±2% in parallel with 20 pF ±3 pF |
| Input Coupling | AC, DC, GND |
| Input Sensitivity Range | 2 mV/div to 5 V/div in a 1-2-5 sequence |
| Vertical Resolution (ADC) | 8 bits |
| DC Gain Accuracy | ±3% for 10 mV/div to 5 V/div, ±4% for 2 mV/div to 5 mV/div |
| Maximum Input Voltage | 300 V RMS CAT II (derated at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above) |
| Channel-to-Channel Isolation | >= 100:1 at 200 MHz |
| Hardware Bandwidth Limits | 20 MHz |
| Lower Bandwidth Limit (AC Coupled) | <10 Hz at BNC |
| Common Mode Rejection Ratio (CMRR) | 100:1 at 60 Hz, 10:1 at 50 MHz |
| Vertical Expansion Range | Allows vertical expansion or compression of active or stopped waveforms |
| Vertical Position Range | ±2 V (2 mV/div to 200 mV/div), ±50 V (>200 mV/div to 5 V/div) |
Horizontal System
| Parameter | Value |
|---|---|
| Maximum Real-Time Sampling Rate | 2.0 GS/s on all channels simultaneously |
| Equivalent-Time Sampling Rate (ETS) | Not supported (Real-time sampling only) |
| Maximum Record Length (Points per Channel) | 2.5k points |
| Time Base Range | 2 ns/div to 50 s/div |
| Time Base Accuracy | 50 ppm |
| Dual Time Base | Yes (allows viewing main and window sweep simultaneously) |
| Horizontal Expansion Range | Allows horizontal expansion or compression of active or stopped waveforms |
| XY Mode Bandwidth | 20 MHz |
| XY Mode Phase Difference | ±3° at 50 kHz |
| Real-time Sample Rate Range | 50 S/s to 2.0 GS/s |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Sources | CH1, CH2, Ext, Ext/5, AC Line |
| Trigger Modes | Auto, Normal, Single Sequence |
| Trigger Coupling | AC, DC, Noise Reject, HF Reject, LF Reject |
| Trigger Holdoff Range | 500 ns to 10 s |
| Edge Triggering | Rising or falling slope (selectable) |
| Pulse Width Triggering | Trigger on pulse width <, >, or = selectable time limit; 33 ns to 10 s |
| Video Triggering | Trigger on NTSC, PAL, or SECAM; field 1, field 2, or line selectable |
| External Trigger Input Impedance | 1 MΩ ±2% in parallel with 20 pF ±3 pF |
| External Trigger Maximum Input Voltage | 300 V RMS CAT II |
| Trigger Sensitivity (Edge, DC Coupled) | Internal: 0.3 div (DC to 10 MHz), 1 div (10 MHz to 200 MHz); External: 100 mV (DC to 10 MHz), 350 mV (10 MHz to 200 MHz) |
| Trigger Level Range | Internal: ±8 divisions from screen center; External: ±1.6 V; External/5: ±8 V |
| Trigger Level Accuracy (typical) | Internal: ±0.2 div; External: ±(6% of setting + 40 mV); External/5: ±(6% of setting + 200 mV) |
| Pulse Width Range | 33 ns to 10 s |
| Pulse Width Resolution | 16.5 ns or 51 ppm, whichever is larger |
| Pulse Width Trigger Conditions | <, >, = (equal), ≠ (not equal) |
| Supported Video Standards | NTSC, PAL, SECAM |
Acquisition System
| Parameter | Value |
|---|---|
| Acquisition Modes | Sample, Peak Detect, Average |
| Peak Detect Mode Min Pulse Width | 12 ns (typical) |
| Averaging Factor Range | 4, 16, 64, 128 |
Waveform Measurements and Analysis
| Parameter | Value |
|---|---|
| Automatic Voltage Measurements | Max, Min, Peak-to-Peak, Mean, Cycle RMS |
| Automatic Time Measurements | Period, Frequency, +Width, -Width, Rise Time, Fall Time |
| Total Automatic Measurements | 11 measurements selectable |
| Basic Math Functions | Addition (+), Subtraction (-) |
| Fast Fourier Transform (FFT) Window Types | Hanning, Flat Top, Rectangular |
| FFT Vertical Scale Options | Linear RMS, dBV RMS |
| Internal Non-Volatile Storage | 4 setups and 2 reference waveforms |
| Cursor Measurements | Amplitude (voltage/current) and Time (delta T, 1/delta T) |
Display and User Interface
| Parameter | Value |
|---|---|
| Waveform Display Formats | YT (voltage vs. time) and XY (channel 1 vs. channel 2) |
| Graticule Options | Grid, Reticule, None |
| Display Graticule Size | 8 × 10 divisions |
| Multi-language User Interface | Yes (10 languages supported) |
| Context-sensitive Help | Yes |
| Probe Check Wizard | Yes (automatically verifies probe attenuation) |
| Autoset Menu | Yes, automatically configures vertical, horizontal, and trigger |
Connectivity and Ports
| Parameter | Value |
|---|---|
| Expansion Slot | Rear-panel slot supports optional TDS2CMA or TDS2MEM modules |
| Optional TDS2CMA Ports | RS-232, Centronics Printer Port, GPIB (IEEE-488) |
| Optional TDS2MEM Ports | CompactFlash Card Slot (supports CompactFlash up to 2 GB), RS-232, Centronics Printer Port |
| Probe Calibration Output | 5.0 V into ≥1 MΩ load at 1 kHz |
Environmental Specifications
| Parameter | Value |
|---|---|
| Storage Temperature Range | -40 to +71 °C |
| Operating Humidity Range | Up to 90% RH at or below +30 °C; up to 60% RH up to +50 °C |
| Operating Altitude | Up to 3,000 m |
| Non-Operating Altitude | Up to 15,000 m |
| Cooling Method | Convection cooling (fanless design) |
Physical Characteristics
| Parameter | Value |
|---|---|
| Form Factor | Benchtop portable |
| Included Probes | Two P2200 200 MHz 10x/1x passive probes |
Compliance and Standards
| Parameter | Value |
|---|---|
| Electromagnetic Compatibility (EMC) | EU EMC Directive, EN61326 Class A, FCC Part 15 Subpart B |
| Calibration Interval | 1 year |
Performance
| Parameter | Value |
|---|---|
| DC Measurement Accuracy (Average Mode) | ±(3% × reading + 0.05 div) for average ≥16 |
Environmental
| Parameter | Value |
|---|---|
| Operating Vibration | 0.31 g RMS from 5 Hz to 500 Hz, 10 minutes on each axis |
| Non-operating Vibration | 2.46 g RMS from 5 Hz to 500 Hz, 10 minutes on each axis |
| Non-operating Shock | 50 g, 11 ms, half-sine |
| Pollution Degree | 2 |
Storage
| Parameter | Value |
|---|---|
| Reference Waveform Storage | 2 internal reference waveforms |
| Setup Storage | 10 internal front-panel setups |
Safety & Compliance
| Parameter | Value |
|---|---|
| Overvoltage Category | CAT II 300 V |
Power Supply
| Parameter | Value |
|---|---|
| Power Consumption | < 30 W |
Triggering
| Parameter | Value |
|---|---|
| Trigger Frequency Counter | 6-digit resolution |
| Trigger Frequency Counter Range | 10 Hz to 200 MHz |
Signal Outputs
| Parameter | Value |
|---|---|
| Probe Compensator Output Frequency | 1 kHz |
| Probe Compensator Output Voltage | 5.0 V into ≥1 MΩ load |
Technology
Digital Storage Oscilloscope (DSO)
Applications
Electronic Design, Troubleshooting, Education, Production Test
Looking for the Tektronix TDS2022?
Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.
Selling this kind of equipment? Sell with us