The Tektronix TDS220 is a highly popular, dual-channel 100 MHz digital storage oscilloscope (DSO) known for its compact form factor, user-friendly interface, and high-speed real-time sample rate of 1.0 GS/s per channel.

Specifications

General

ParameterValue
InterfacesNone standard; RS-232, GPIB, and Centronics parallel ports are available via optional TDS2CM or TDS2MM communication modules
DisplayHigh-contrast monochrome LCD with backlighting, 320 × 240 pixels, contrast adjustment

Physical

ParameterValue
Operating Temperature0 to +50 °C
Dimensions (W×H×D)304.8 mm × 151.4 mm × 120.7 mm
Weight1.5 kg (instrument only)

Vertical System

ParameterValue
Number of Channels2
Analog Bandwidth100 MHz
Calculated Rise Time3.5 ns
Input CouplingAC, DC, GND
Input Sensitivity Range2 mV/div to 5 V/div in a 1-2-5 sequence
Vertical Resolution8 bits (1 part in 256)
DC Gain Accuracy±3% for Sample or Average acquisition mode
Maximum Input Voltage300 V RMS CAT II (derated above 100 kHz to 13 V peak AC at 3 MHz and above)
Position Range±2 V for 2 mV/div to 200 mV/div; ±50 V for 500 mV/div to 5 V/div
Analog Bandwidth Limit20 MHz

Horizontal System

ParameterValue
Time Base Range5 ns/div to 5 s/div in a 1-2-5 sequence

Trigger System

ParameterValue
Trigger TypesEdge (rising or falling), Video
Trigger SourcesCH1, CH2, EXT, EXT/5, AC Line
External Trigger Input Impedance1 MΩ in parallel with 20 pF
External Trigger Maximum Input300 V RMS CAT II

Acquisition System

ParameterValue
Max Real-Time Sample Rate1.0 GS/s per channel
Record Length2500 samples per channel
Acquisition ModesSample, Peak Detect, Average
Peak Detect Glitch Capture WidthCaptures high-frequency and random glitches as narrow as 10 ns
Average Acquisition ModeSelectable from 4, 16, 64, or 128 waveforms
Non-Volatile Waveform StorageTwo 2500-point reference waveforms
Non-Volatile Setup Storage5 instrument setups

Waveform Measurements and Math

ParameterValue
Automatic MeasurementsPeriod, Frequency, Cycle RMS, Mean, Peak-to-Peak
Cursor MeasurementsVoltage difference (ΔV) and Time difference (Δt, 1/Δt)
Waveform Math OperationsCH1 + CH2, CH1 - CH2
FFT AnalysisAvailable via optional TDS2MM extension module (Hanning, Flat Top, Rectangular windows)

Display and User Interface

ParameterValue
Display Resolution320 × 240 pixels
Multi-Language User InterfaceEnglish, French, German, Italian, Spanish, Portuguese, Japanese, Korean, Simplified Chinese, Traditional Chinese
Autoset MenuAutomatically adjusts vertical, horizontal, and trigger controls for a stable display

Connectivity and Expansion

ParameterValue
Extension Module InterfaceRear slot for optional communications and feature-expansion modules
Optional ModulesTDS2CM (adds RS-232, GPIB, Centronics), TDS2MM (adds RS-232, GPIB, Centronics, FFT, Rise/Fall times)
Probe Compensator OutputApproximately 5 V into ≥1 MΩ load, 1 kHz frequency

Physical Characteristics

ParameterValue
Height151.4 mm
Width304.8 mm
Depth120.7 mm

Environmental and Safety

ParameterValue
Storage Temperature-20 °C to +60 °C
Operating AltitudeUp to 3,000 meters
Non-Operating AltitudeUp to 15,000 meters
Safety CertificationUL3111-1, CSA1010.1, EN61010-1
Electro-Magnetic CompatibilityMeets directive 89/336/EEC, EN 55011 Class A, EN 50082-1

Trigger

ParameterValue
Trigger Level Range (Internal)±8 divisions from center of screen
Trigger Level Range (External)±1.6 V
Trigger Level Range (External/10)±16 V
Trigger Holdoff Range500 ns to 10 s

Measurements

ParameterValue
Automatic Measurement TypesPeriod, Frequency, Cycle RMS, Mean, Peak-to-Peak
Cursor Measurement TypesVoltage (ΔV), Time (ΔT), Frequency (1/ΔT)

General Specifications

ParameterValue
Probe Compensator Output VoltageApproximately 5 V into ≥ 1 MΩ load
Probe Compensator Output FrequencyApproximately 1 kHz
Calibration Interval1 year (recommended)

Safety & Compliance

ParameterValue
Overvoltage Category (Inputs)CAT II 300 Vrms
Overvoltage Category (Mains)CAT II
Pollution DegreePollution Degree 2 (indoor use only)

Display

ParameterValue
Active Display Area115 mm × 86 mm (4.5 in × 3.4 in)

Technology

Digital Storage Oscilloscope (DSO)

Applications

Electronic design, debugging, troubleshooting, education, service, and repair