The Tektronix TDS3012C is a 100 MHz, 2-channel Digital Phosphor Oscilloscope (DPO) from the popular TDS3000C series. Known for its compact design, optional battery capability, and rapid 3,600 wfms/s continuous waveform capture rate, it provides reliable performance for engineering, manufacturing, and service environments.

Specifications

General

ParameterValue
InterfacesUSB Host (front panel), USB Device (rear panel), LAN (10Base-T); GPIB, RS-232, VGA video output (optional TDS3GV module)
Display6.5-inch (165.1 mm) active-matrix color TFT, 640 × 480 pixels
Standard Warranty3 years (excluding probes)

Physical

ParameterValue
Power Supply100 V to 240 V AC ±10%, 47 Hz to 440 Hz (100 V to 120 V), 47 Hz to 63 Hz (115 V to 240 V), 75 W maximum
Operating Temperature0 °C to +50 °C
Dimensions (W×H×D)375 mm × 176 mm × 149 mm
Weight3.2 kg

Vertical System (Analog Channels)

ParameterValue
Number of Channels2
Analog Bandwidth (-3 dB)100 MHz
Calculated Rise Time3.5 ns
Hardware Bandwidth Limits20 MHz
Input CouplingAC, DC, GND
Input Impedance1 MΩ ±1% in parallel with 13 pF ±2 pF, or 50 Ω ±1%
Input Sensitivity Range (1 MΩ)1 mV/div to 10 V/div
Input Sensitivity Range (50 Ω)1 mV/div to 1 V/div
Vertical Resolution9 bits
Maximum Input Voltage (1 MΩ)150 V RMS CAT I
Maximum Input Voltage (50 Ω)5 V RMS
Channel-to-Channel Isolation≥100:1 at 100 MHz

Horizontal System

ParameterValue
Time Base Range4 ns/div to 10 s/div
Time Base Accuracy±20 ppm

Acquisition System

ParameterValue
Max Real-Time Sample Rate1.25 GS/s
Record Length (Max Points)10 k points
Acquisition ModesSample, Peak Detect, Envelope, Average, Single Sequence
Peak Detect Resolution9.3 ns
Continuous Waveform Capture Rate3,600 wfms/s

Trigger System

ParameterValue
Trigger ModesAuto, Normal, Single
Trigger SourceCH1, CH2, AC Line, External, External/10
Trigger CouplingAC, DC, Noise Reject, HF Reject, LF Reject

Waveform Measurements & Analysis

ParameterValue
Automatic Measurements25
Fast Fourier Transform (FFT)Standard
Waveform Math OperationsAdd, Subtract, Multiply, Divide, FFT
Waveform Anomaly DetectionWaveAlert®
Optional Analysis ModulesAdvanced Analysis, Limit Testing, Telecom Mask Testing, Extended Video, 601 Serial Digital Video

Display System

ParameterValue
Display TypeVGA Active Matrix Color TFT
Display Size6.5 in (165.1 mm) diagonal
Display Resolution640 × 480 pixels
Graticule TypesFull, Grid, Crosshair, Frame
Display FormatsYT, XY, Zoom
Waveform Display StylesVectors, Dots, Variable Persistence, Infinite Persistence

Input/Output Ports

ParameterValue
USB Host Port1x USB 2.0 (front panel)
Ethernet Port (LAN)1x RJ-45 (10Base-T)
VGA Video Output1x DB-15 female (requires optional TDS3GV module)
External Trigger InputBNC, 1 MΩ
Probe Compensator Output5 V into ≥1 MΩ, 1 kHz

Power and Battery System

ParameterValue
Line Voltage Range100 V to 240 V AC ±10%
Line Frequency Range47 Hz to 440 Hz (100–120 V), 47 Hz to 63 Hz (115–240 V)
Power Consumption75 W maximum
Optional Battery OperationTDS3BATC lithium-ion battery pack
Battery Life (Typical)Up to 3 hours

Environmental and Safety

ParameterValue
Non-Operating Temperature-20 °C to +60 °C
Operating HumidityUp to 95% RH at or below +30 °C, up to 45% RH up to +50 °C
Operating AltitudeUp to 3,000 m
Non-Operating AltitudeUp to 15,000 m

Physical Characteristics

ParameterValue
Width375 mm
Height176 mm
Depth149 mm
Weight (Instrument Only)3.2 kg
Weight (with Battery)4.5 kg

Trigger

ParameterValue
Trigger Level RangeInternal: ±8 divisions from screen center; External: ±1.2 V; External /10: ±12 V
Trigger Holdoff Range250 ns to 8 s
External Trigger Input Impedance1 MΩ in parallel with 17 pF
Internal Edge Trigger Sensitivity0.35 div from DC to 50 MHz, increasing to 1 div at 100 MHz
External Edge Trigger Sensitivity150 mV from DC to 50 MHz, increasing to 500 mV at 100 MHz
External /10 Edge Trigger Sensitivity1.5 V from DC to 50 MHz, increasing to 5 V at 100 MHz

Vertical System

ParameterValue
Channel-to-Channel Deskew Range±10 ns
Common Mode Rejection Ratio100:1 at 60 Hz, 10:1 at 50 MHz (with equal Volts/div settings)
Input Capacitance (1 MΩ)13 pF ±2 pF

Compliance

ParameterValue
Safety ComplianceUL61010B-1, CSA C22.2 No. 1010.1, EN61010-1
EMC ComplianceEN61326 Class A, FCC Part 15 Subpart B Class A, AS/NZS 2064

Environmental

ParameterValue
Operating Shock30 g, 11 ms half-sine
Non-Operating Shock50 g, 11 ms half-sine
Operating Vibration0.31 g RMS from 5 Hz to 500 Hz
Non-Operating Vibration2.46 g RMS from 5 Hz to 500 Hz
Cooling Clearance RequirementAt least 51 mm (2 in.) on left side and rear of instrument

Accessories

ParameterValue
Standard Probes ShippedTwo P3010 100 MHz 10X passive probes

Storage

ParameterValue
Internal Setup Storage10 setups
Internal Waveform Storage4 waveforms

User Interface

ParameterValue
Autoset FunctionSingle-button automatic setup of vertical, horizontal, and trigger systems

Interfaces

ParameterValue
GPIB InterfaceSupported via optional TEK-USB-488 Adapter connected to USB device port

Technology

Digital Phosphor Oscilloscope (DPO)

Applications

Electronic Design, Debugging, Manufacturing Test, Service and Repair, Education