The Tektronix TDS320 is a 100 MHz, two-channel digital storage oscilloscope designed for electronic testing and troubleshooting. It offers a 500 MS/s real-time sampling rate on each channel, an 8-bit vertical resolution, and advanced horizontal and vertical systems for accurate waveform measurement and analysis. With the Option 14 computer interface, it supports comprehensive GPIB, RS-232, and hardcopy capabilities.
Specifications
General
| Parameter | Value |
|---|---|
| Interfaces | GPIB (IEEE-488.2), RS-232, Centronics-type parallel (with Option 14) |
| Display | Monochrome CRT |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 90 to 250 VAC continuous, 45 Hz to 440 Hz, < 65 W |
| Operating Temperature | 0 to +50 °C |
| Dimensions (W×H×D) | 362 mm × 165 mm × 472 mm |
| Weight | 6.5 kg |
Vertical System
| Parameter | Value |
|---|---|
| Analog Bandwidth (-3 dB) | 100 MHz |
| Number of Input Channels | 2 |
| Input Sensitivity Range | 2 mV/div to 10 V/div |
| Vertical Resolution (ADC) | 8 bits |
| Vertical Divisions | 256 levels over 10.24 vertical divisions |
| Channel Invert Function | Supported on each channel |
| Input Impedance | 1 MΩ ±1% in parallel with 20 pF ±2 pF |
| Input Coupling | AC, DC, GND |
| Maximum Input Voltage | 400 V (DC + peak AC) |
| DC Gain Accuracy | ±2% |
Horizontal System
| Parameter | Value |
|---|---|
| Maximum Real-Time Sampling Rate | 500 MS/s on each channel |
| Maximum Record Length | 1000 sample points per channel |
| Time Base Range | 5 ns/div to 5 s/div |
| Time Base System | Main and Delayed |
| Time Base Accuracy | ±0.01% |
| Waveform Interpolation | Sin(x)/x |
Trigger System
| Parameter | Value |
|---|---|
| Trigger Sources/Types | Edge, Video |
| Trigger System Capabilities | Main Only |
| Trigger Coupling | AC, DC, HF Reject, LF Reject, Noise Reject |
| External Trigger Input Impedance | Not supported |
| External Trigger Max Input Voltage | Not supported |
Waveform Measurements and Analysis
| Parameter | Value |
|---|---|
| Automatic Measurements | Supported |
Connectivity and Ports
| Parameter | Value |
|---|---|
| GPIB Interface Option | IEEE-488.2 programmable (Option 14) |
| RS-232 Interface Option | RS-232 programmable (Option 14) |
| Video Output Port | Not supported |
| Hardcopy Port | Centronics-type parallel or RS-232 (Option 14) |
| Graphics Interface File Formats | Interleaf (.img), TIFF, PCX, BMP, Encapsulated Postscript (EPS) (Option 14) |
| Supported Printer Formats | HP ThinkJet, DeskJet, LaserJet, Epson (9 & 24 Pin), Seiko DPU 411/II, DPU 412 (Option 14) |
| Printer DC Power Output | Provided for Seiko printers (Option 14) |
Acquisition System
| Parameter | Value |
|---|---|
| Acquisition Modes | Sample, Envelope, Average, Peak Detect |
| Glitch Capture (Peak Detect) | 10 ns |
Measurement & Math
| Parameter | Value |
|---|---|
| Waveform Math Functions | CH1 + CH2, CH1 - CH2, CH2 - CH1 |
Storage & Save
| Parameter | Value |
|---|---|
| Storage (Reference Waveforms) | Two 1k point reference waveforms (Ref1, Ref2) |
Auxiliary Outputs
| Parameter | Value |
|---|---|
| Probe Compensator Output Voltage | 5.0 V into ≥1 MΩ load |
| Probe Compensator Output Frequency | 1.2 kHz |
Power & Environmental
| Parameter | Value |
|---|---|
| Power Consumption | ≤ 65 W (90 VA) |
| Operating Altitude | 4,570 m (15,000 ft) |
| Non-Operating Altitude | 12,190 m (40,000 ft) |
Setup & Waveform Storage
| Parameter | Value |
|---|---|
| Setup Storage Capacity | 10 front-panel setups |
Certifications & Compliance
| Parameter | Value |
|---|---|
| Safety Compliance | UL 1244, CSA C22.2 No. 231, EN 61010-1 |
Technology
Digital Storage Oscilloscope (DSO)
Applications
Electronic design, debugging, troubleshooting, and education