The Tektronix TPS2012 is a 100 MHz, 2-channel Digital Storage Oscilloscope in the TPS2000 Series. Designed for engineers and technicians needing to make floating and differential measurements, it features IsolatedChannel technology for up to 600 VRMS CAT II or 300 VRMS CAT III isolation between channels and to ground. Compact and battery-powered, the TPS2012 offers high-performance digital real-time sampling and versatile analysis capabilities from the bench to the field.

Specifications

General

ParameterValue
InterfacesRS-232 (9-pin D-sub male), Centronics parallel printer port, CompactFlash Type I card slot
Display5.7 inch (145 mm) color active-matrix TFT LCD, 320 × 240 pixels
Recommended Calibration Interval1 year
Standard Warranty3 years

Physical

ParameterValue
Power SupplyExternal AC adapter (100 to 240 VAC, 50/60 Hz to 15 VDC output) or up to 2 rechargeable Lithium-Ion batteries
Operating Temperature0 °C to +50 °C
Dimensions (W×H×D)336.0 mm × 161.0 mm × 130.0 mm
Weight2.7 kg (instrument without battery)

Vertical System

ParameterValue
Analog Bandwidth (-3 dB)100 MHz
Number of Channels2 isolated channels
Vertical Resolution8 bits
Input Impedance1 MΩ ±2% in parallel with 20 pF ±3 pF
Input CouplingAC, DC, GND
Input Sensitivity Range2 mV/div to 5 V/div
DC Gain Accuracy±3%
Maximum Input Voltage (BNC Signal to Shell)300 VRMS CAT II
Maximum Float Voltage (BNC Shell to Earth Ground)600 VRMS CAT II, 300 VRMS CAT III
Hardware Bandwidth Limits20 MHz
Common Mode Rejection Ratio (CMRR)100:1 at 60 Hz, 20:1 at 50 MHz (with P2220 probe)
Channel-to-Channel Isolation≥100:1 at 50 MHz

Horizontal System

ParameterValue
Maximum Real-Time Sampling Rate1.0 GS/s per channel
Maximum Record Length2500 points per channel
Time Base Range5 ns/div to 50 s/div
Time Base Accuracy±50 ppm

Trigger System

ParameterValue
Trigger SourcesChannel 1, Channel 2, Ext (Isolated External Trigger Input)
Trigger ModesAuto, Normal, Single Sequence
Trigger CouplingAC, DC, Noise Reject, HF Reject, LF Reject
Edge TriggeringRising, falling
Video TriggeringNTSC, PAL, SECAM
Pulse Width TriggeringTrigger on pulse width <, >, =, or ≠; 33 ns to 10 s
Internal Trigger Sensitivity0.35 div (DC to 10 MHz), 1.0 div (10 MHz to 100 MHz)
External Trigger Sensitivity200 mV (DC to 10 MHz), 375 mV (10 MHz to 100 MHz)

Acquisition System

ParameterValue
Peak Detect Mode Min Pulse Width12 ns
Averaging Factor Range4 to 128

Waveform Measurements and Analysis

ParameterValue
Automatic Measurements11 measurements (Period, Frequency, +Width, -Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, Cycle RMS)
Fast Fourier Transform (FFT)Standard
FFT Window TypesHanning, Flat Top, Rectangular

Display and User Interface

ParameterValue
Graticule OptionsGrid, Crosshair, Frame

Connectivity and Ports

ParameterValue
Isolated External Trigger InputYes, 1 isolated channel

Power Specifications

ParameterValue
Battery TypeTPSBAT Lithium-Ion battery pack
Number of Batteries SupportedUp to 2 battery packs (hot-swappable)
Battery Operation TimeUp to 4 hours per battery (8 hours total with 2 batteries)

Environmental Specifications

ParameterValue
Storage Temperature-40 °C to +71 °C
Operating Humidity (≤ +30 °C)Up to 90% RH
Operating Humidity (+31 °C to +50 °C)Up to 60% RH
Operating AltitudeUp to 3,000 meters (9,843 feet)
Cooling MethodConvection cooled (fanless)

Physical Characteristics

ParameterValue
Weight (without batteries)2.7 kg (6.0 lbs)
Weight (with one battery)3.2 kg (7.0 lbs)
Weight (with two batteries)3.7 kg (8.1 lbs)

Performance

ParameterValue
Rise Time3.5 ns

Storage & Memory

ParameterValue
CompactFlash Storage CapacityType I, up to 2 GB
Reference Waveform Storage2 reference waveforms
Setup Storage10 front-panel setups

Measurements

ParameterValue
Cursor MeasurementsAmplitude, Time

Waveform Measurements

ParameterValue
Waveform Math OperationsAddition, Subtraction, Multiplication, FFT
FFT Resolution2048 points

Signal Outputs

ParameterValue
Probe Compensator Output VoltageApprox. 5 V into ≥1 MΩ load
Probe Compensator Output Frequency1 kHz

Display System

ParameterValue
Display FormatsYT and XY

Ease of Use

ParameterValue
Autoset FunctionSingle-button automatic setup of vertical, horizontal, and trigger systems, with undo function

Compliance

ParameterValue
Overvoltage CategoryCAT II 300 V (BNC Signal to Common), CAT II 600 V / CAT III 300 V (Common to Earth Ground)
Safety CertificationUL61010-1, CAN/CSA-C22.2 No. 61010-1, EN61010-1
Electromagnetic CompatibilityMeets Directive 89/336/EEC, EN61326 Class A

Software & Options

ParameterValue
Optional Power Analysis SoftwareTPS2PWR1

Technology

Digital Storage

Applications

Floating and differential measurements, industrial power system debugging, motor drive design, and field troubleshooting

Looking for the Tektronix TPS2012?

Tell us what you need. We're building a network of equipment dealers and will get back to you if we can help source it — no obligation.

Selling this kind of equipment? Sell with us