The Rohde & Schwarz FSWP8 is a high-performance Phase Noise Analyzer and VCO Tester designed for extremely fast and highly sensitive phase noise measurements in the 1 MHz to 8 GHz frequency range. Combining phase noise and signal source analysis in one instrument, it supports cross-correlation to achieve unparalleled sensitivity and offers an optional full-featured signal and spectrum analyzer mode to cover diverse test and measurement needs.

Specifications

General

ParameterValue
InterfacesLAN (10/100/1000Base-T), USB 2.0 (Host & Device), USB 3.0, DVI-D, DisplayPort, GPIB (optional)
Display12.1-inch WXGA color TFT touchscreen (1280 × 800 pixels)
Standard Measurement ModesPhase Noise, AM Noise, VCO Characterization
Optional Measurement ModesSignal and Spectrum Analyzer (R&S FSWP-B3), Pulsed Phase Noise (R&S FSWP-K4), Transient Analysis (R&S FSWP-K60)
Command SetSCPI

Physical

ParameterValue
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz / 400 Hz, 120 W typical power consumption
Dimensions (W×H×D)462 mm × 240 mm × 504 mm
Weight18.6

RF Input & Frequency Characteristics

ParameterValue
RF Input Frequency Range1 MHz to 8 GHz
Input Impedance50 Ω
Maximum Safe RF Input Power+30 dBm
Maximum Input DC Voltage0 V
RF Input ConnectorN female
Internal Local Oscillator Frequency Range1 MHz to 8 GHz

Phase Noise Measurement

ParameterValue
Offset Frequency Range10 mHz to 30 MHz
Phase Noise Measurement Accuracy±1.5 dB (nominal)
AM Noise Measurement Offset Range10 mHz to 30 MHz

Internal Reference

ParameterValue
Reference Output Frequency10 MHz
External Reference Input Frequency Range1 MHz to 20 MHz (in 1 Hz steps)
External Reference Input Level-10 dBm to +15 dBm

DC Bias and Signal Sources

ParameterValue
Tuning Voltage Range (VPTC)-10 V to +28 V (with R&S FSWP-B60/B61 option)
Tuning Voltage Resolution0.1 mV (with R&S FSWP-B60/B61 option)
Tuning Voltage Maximum Current20 mA (with R&S FSWP-B60/B61 option)
Auxiliary Voltage Range (VAUX)-10 V to +15 V (with R&S FSWP-B60/B61 option)
Auxiliary Voltage Resolution0.1 mV (with R&S FSWP-B60/B61 option)
Auxiliary Voltage Maximum Current250 mA (with R&S FSWP-B60/B61 option)
Internal RF Stimulus Source Frequency Range1 MHz to 8 GHz (with R&S FSWP-B64 option)

Display & UI

ParameterValue
Measurement Screen FormatsMultiView (simultaneous display of multiple measurement tasks)
Built-In Operating SystemWindows

Connectivity & Interfaces

ParameterValue
LAN Port Speed10/100/1000Base-T (RJ-45)
USB Host Ports (Front)4 × USB 2.0
USB Host Ports (Rear)2 × USB 3.0, 2 × USB 2.0
USB Device Port1 × USB 2.0 Type B
External Monitor Outputs1 × DVI-D, 1 × DisplayPort
External Trigger InputBNC female
GPIB Interface PortOption R&S FSW-B10 (IEEE 488)
Noise Source Control PortBNC female, 28 V DC

Power Requirements

ParameterValue
Input Voltage Range100 V to 240 V AC
Input AC Frequency50 Hz to 60 Hz / 400 Hz
Maximum Power Consumption220 W (with all options)
Typical Power Consumption120 W

Physical and Environmental Specifications

ParameterValue
Rack Unit Height5 RU
Storage Temperature Range-40 °C to +70 °C
Operating Humidity Range95% relative humidity (non-condensing)
Maximum Operating Altitude4600 m (15,000 feet)

Compliance and Standards

ParameterValue
Calibration Interval1 year

Display

ParameterValue
Display Size12.1 in
Display Resolution1280 × 800 pixels

Reference Frequency

ParameterValue
Standard Reference Frequency Aging1 × 10^-6 per year
Standard Reference Frequency Temperature Stability1 × 10^-6 (0 °C to +50 °C)
Precision Reference Frequency Option Aging3 × 10^-8 per year (with R&S FSWP-B4 option)
Precision Reference Frequency Option Temperature Stability5 × 10^-10 (0 °C to +50 °C, with R&S FSWP-B4 option)

Spectrum Analyzer Option

ParameterValue
Spectrum Analyzer Option Frequency Range10 Hz to 8 GHz (with R&S FSWP-B1 option)
Spectrum Analyzer Option Resolution Bandwidth (RBW)1 Hz to 10 MHz (with R&S FSWP-B1 option)
Spectrum Analyzer Preamplifier Option Frequency Range100 kHz to 8 GHz (with R&S FSWP-B24 option)
Spectrum Analyzer Option DANL (1 GHz)<-152 dBm (1 Hz RBW, preamplifier off, typical)
Spectrum Analyzer Option TOI (1 GHz)> 20 dBm (typical +25 dBm)

VCO Tester Source & Measurement

ParameterValue
Tuning Voltage (VPTC) Setting Uncertainty±(0.05% of setting + 1.5 mV)
Auxiliary Voltage (VAUX) Setting Uncertainty±(0.05% of setting + 1.5 mV)

RF Input

ParameterValue
RF Input VSWR< 1.2 (10 MHz to 3 GHz, RF attenuation ≥ 10 dB), < 1.5 (3 GHz to 8 GHz, RF attenuation ≥ 10 dB)

AM Noise Measurement

ParameterValue
AM Noise Measurement Carrier Frequency Range1 MHz to 8 GHz

Internal RF Stimulus Source Option

ParameterValue
Internal RF Stimulus Source Output Power Range-50 dBm to +10 dBm (range is -110 dBm to +10 dB with optional step attenuator)

Reference Signals

ParameterValue
External Reference Input Impedance50 Ω
External Reference Output Impedance50 Ω

Technology

Digital Cross-Correlation Signal Analysis

Applications

Phase noise measurement, VCO characterization, AM noise analysis, and additive phase noise testing.