The Rohde & Schwarz SMHU is a high-precision, low-noise RF signal generator covering the frequency range from 100 kHz to 4.32 GHz. Designed for demanding applications in laboratory research, development, and high-throughput production environments, it provides exceptional spectral purity, fast frequency switching, and versatile modulation capabilities including AM, FM, Phase, and high-performance Pulse modulation.
Specifications
Frequency
| Parameter | Value |
|---|---|
| Minimum Frequency | 100 kHz |
| Maximum Frequency | 4.32 GHz |
| Frequency Resolution | 0.1 Hz |
Output
| Parameter | Value |
|---|---|
| Maximum Output Power | +13 dBm |
| Output Power Range | -140 dBm to +13 dBm (overrange up to +19 dBm) |
| Phase Noise | <-140 dBc/Hz (typical -143 dBc/Hz) at 20 kHz offset (1 GHz carrier) |
Modulation
| Parameter | Value |
|---|---|
| Modulation | AM, FM, Phase (̆M), Pulse |
| Sweep Modes | Frequency, Level, Modulation Frequency (linear/logarithmic) |
| External AM Frequency Range | DC to 50 kHz (3 dB bandwidth) |
| External FM Frequency Range | DC to 100 kHz (standard), DC to 1 MHz (broadband FM mode) |
| External Phase Modulation Frequency Range | 20 Hz to 20 kHz |
| Internal Modulation Generators | 2 independent synthesizers (standard) |
General
| Parameter | Value |
|---|---|
| Reference Oscillator | 10 MHz OCXO |
| Interfaces | GPIB (IEEE-488) |
| Display | LCD |
| Memory Storage Locations | 100 complete instrument setups |
| Warm-up Time | approx. 10 min |
Physical
| Parameter | Value |
|---|---|
| Power Supply | 100 V / 115 V / 120 V / 200 V / 220 V / 240 V AC (±10%), 47 Hz to 440 Hz |
| Dimensions (W×H×D) | 435 mm × 236 mm × 570 mm |
| Weight | 31 kg |
| Rack Units | 5 RU |
Frequency Specifications
| Parameter | Value |
|---|---|
| Frequency Settling Time | < 1 ms |
| Reference Frequency | 10 MHz |
| Aging Rate | < 1 x 10^-9 / day (after warm-up) |
| Temperature Stability | < 2 x 10^-9 / °C |
Level (Amplitude) Specifications
| Parameter | Value |
|---|---|
| Output Impedance | 50 Ω |
| Level Resolution | 0.1 dB |
Spectral Purity
| Parameter | Value |
|---|---|
| Harmonics | <-30 dBc (for output levels < +10 dBm) |
| Non-Harmonic Spurious | <-100 dBc (for carrier frequencies f < 1 GHz, offset > 10 kHz) |
| Residual FM | < 1 Hz (at 1 GHz, 0.3 kHz to 3 kHz bandwidth) |
| Subharmonics (1000 to 2000 MHz) | <-76 dBc |
| Subharmonics (2000 to 4320 MHz) | <-70 dBc |
| Non-Harmonic Spurious (< 1000 MHz) | <-100 dBc (at > 10 kHz offset) |
| Non-Harmonic Spurious (1000 to 2000 MHz) | <-94 dBc (at > 10 kHz offset) |
| Non-Harmonic Spurious (2000 to 4320 MHz) | <-88 dBc (at > 10 kHz offset) |
Amplitude Modulation (AM)
| Parameter | Value |
|---|---|
| AM Modulation Depth Range | 0% to 99% |
| AM Resolution | 0.1% |
| AM Distortion | < 1% (at 30% AM, f_mod = 1 kHz) |
| AM Frequency Response | DC to 50 kHz (3 dB bandwidth) |
Frequency Modulation (FM)
| Parameter | Value |
|---|---|
| FM Frequency Response | DC to 100 kHz (3 dB bandwidth) |
Pulse Modulation
| Parameter | Value |
|---|---|
| Pulse Modulation On/Off Ratio | > 80 dB |
| Pulse Modulation Rise/Fall Time | < 20 ns |
Internal Modulation Source
| Parameter | Value |
|---|---|
| Internal Waveform Types | Sine, Square, Triangular, Sawtooth |
| Internal Modulation Frequency Range | 0.1 Hz to 100 kHz |
| Internal Modulation Frequency Resolution | 0.1 Hz |
General and Environmental
| Parameter | Value |
|---|---|
| AC Line Voltage Range | 100 V to 240 V AC (±10%) |
| AC Line Frequency Range | 47 Hz to 440 Hz |
| Power Consumption | 230 VA |
| Operating Temperature Range | 0 °C to 55 °C |
Rear Panel and Auxiliary Interfaces
| Parameter | Value |
|---|---|
| RF Output Connector | N female (front panel) |
RF Output
| Parameter | Value |
|---|---|
| Level Accuracy | ±1 dB (for output levels > -127 dBm) |
Reference Frequency
| Parameter | Value |
|---|---|
| External Reference Input | 5 MHz or 10 MHz |
| Reference Output | 10 MHz |
Overload Protection
| Parameter | Value |
|---|---|
| Reverse Power Protection | 50 W (up to 1000 MHz), 25 W (1000 to 4320 MHz) from 50 Ω source |
| Max. DC Overload Voltage | 35 V |
Modulation Options
| Parameter | Value |
|---|---|
| I/Q Modulation Carrier Frequency Range | 1.5 MHz to 4320 MHz (with Option SMHU-B1) |
| I/Q Modulation Bandwidth (3 dB) | DC to 5 MHz (with Option SMHU-B1) |
Technology
Synthesized RF Signal Generator
Applications
RF testing, local oscillator substitution, component and receiver testing